Graduated measuring ranges 0.01 – 10 mg/l
Chlorine dioxide measurement with DULCOTEST® sensors offers precise, reliable, application-specific measured values in real time for all oxidation and disinfection tasks.
Field of application
Product range CDE: Potable, industrial and process water without surfactants with slow temperature changes
Product range CDP: Bottle-washing systems and other applications with addition of surfactants and rapid temperature changes
Product range CDR: Polluted industrial / process water, surfactant-containing, cooling water, irrigation water, waste water with little pollution, hot water
Technical Details
Sensor type CDE 2-mA
- Measuring ranges 0.01-10 mg/l
- Temperature 5 – 45 ºC
- Pressure max. 1.0 bar
- pH range 4.0…11.0
- Response time 120 s
Sensor type CDP 1-mA
- Measuring ranges 0.02-2.00 mg/l
- Temperature 10 - 45 ºC
- Pressure max. 3.0 bar
- pH range 5.5…10.5
- Response time 60 s
Sensor type CDR 1-mA
- Measuring ranges 0.01-10.0 mg/l
- Temperature 1 - 55 ºC
- Pressure max. 3.0 bar
- pH range 1.0…10.0
- Response time 180 s